MODEL 600-9/9500 - TECHNICAL DATA
Gas (2) Sizing Guide
ENGINEERING SPECIFICATIONS
 
Accuracy: +/- .50% of Full Scale, [or 2.0% of reading whichever is better. Full Scale over the calibrated range. [Process fluid temperature span +/- 50°F]
 
Repeatability: +/- 0.2 % of Rate
 
Response Time:
Gases: 1 to 2 sec. Typical
Liquids: Less than 500ms
 
Pressure Rating:
0 - 1200 psig (Std)
0 - 10000 psig Dependent on meter size
 
Mass Flow Rates:
Gases, Liquids: Consult factory for your application.
Note: maximum determined by pipe capacity
 
Flow Range: 1OO:1 Capability
 
Wetted Materials: 316SS (STD). Consult factory for other material requirements
 
Pressure Drop Typically: Same as equivalent size and length tube or pipe
 
Mounting: Flange; Tube Fittings, Sanitary, Welded. Many others available
 
MICROPROCESSOR SPECIFICATIONS:
 
EEPROM Memory: 64 Byte
 
Displays: 8 digit Flowrate and Totalizer (Optional)
 
Communicator:
 
Interface: RS-232 via Ribbon Connector
Baud Rate: 9600
 
Adjustable Variables with RS-232 Communicator:
Zero Cutoff; Zero Offset
Digital Filter Factor (damping 0 to 32 minutes)
Flow Full Scale
Flow Rate Decimal Point
Flow Totalizer Decimal Point
Flow Totalizer/Flowrate Ratio
 
Power: 120/240 vat, 50/60 Hz, 20 Watts
24VDC, 1 Amp (factory set)
 
Electrical Connections: 3/4 ” FNPT Power & Signal
 
Enclosure: Cast Aluminum, Epoxy Painted
 
ENVIRONMENTAL RATINGS
 
Housing: Explosion Proof, Class 1, Div. 1, Groups, B, C, D Weatherproof, Nema 4
 
Ambient Temperature: -25°C to 60°C ( -13°F to 140°F)
 
Storage Temperature: -40°C to 70°C ( -40°F to 158°F )
 
Mounting:
Integral to the Flow Transducer
Remote to 4000 ft., Wall or Frame mount
 
Signal Cable for Remote Design: 18 gauge, 2 pair with cable shield. Maximum distance, 4000 ft.
 
MODEL NUMBERING SYSTEM:
Base Model: = 9500
In-Line: = 600-9/9500
Temperature Output: = 600-9/9500
Electronics Mounting
I = Integral Mount
R = In-Line , Remote Mount
 
Size: 1/4'’ to 12”
 
Material:
316SS = STD
Other = Consult Factory
 
DESIGN INTERIOR FINISH TYPE:
IND = Industrial
SAN= Sanitary
UHP=Ultra High Pure
 
Process Connection:
STB = Stub End for User Fitting.(STD.)
TBE = Tube Fitting (STD.)
WLD = Weld End (STD.)
NPT = MNPT (STD.)
FLG-150 = 150# RF Flange (STD.)
FLG-OTHER = Consult Factory
OTHER = Consult Factory
 
Power:
24 = 22-26 VDC, 1 AMP. 4 WIRE
120 = 120VAC, 50/60 HZ (STD.)
240 = 240VAC 50/60 HZ
 
Output Signals:
(Select One**)
0/5 = 0-5 VDC**
0/10 = 0-10 VDC**
4/20 = 4-20 mADC** into 750 ohms, Isolated std, Self Powered
4/20T = Opt Process Fluid Temp. 4-20mA Self Powered
Other = Consult Factory
 
Display: (Optional)
ND = No Display
D = Flowrate Display, 8 Digit
8T = Totalizer
D/8T = Flowrate and Total Display, 8 digit
 
Signal Cable for Remote Design: SC25 = 25 FT Supplied with remote configuration